    Specifications:

    ECG:
    5 lead or 12 lead selectable
    Input: 5 lead; RA, LT, RL, LL, V or R, L, N, F, C
    Lead selection: I, II, III, aVR, aVL, aVF, Vx, CAL
    Gain Selection: ×25, ×0.5, ×1, ×2, auto
    Sweep speed: 12.5mm/s, 25mm/s, 50mm/s
    Accuracy: ±1bpm or ±1% whichever is grater
    Protection: Withstand 4000V AC/50Hz voltage in isolation against electrosurgical and defibrillation
    Calibration signal: 1mV±5%
    Operation modes, Diagnostic, Monitor, Surgery
    Heart rate range: Adult: 15-3000bpm; Pediatric/Neonatal: 15-350bpm
    Measurement range: -2.0mV to 2.0mV
    ST detection: YES,
    Arrhythmia analysis: YES
    Alarm: YES, audible and visual alarm, alarm events recallable.

    NIBP:
    Method: Oscillometry
    Mode: Manual/Auto/STAT
    Measurement Range: adult 10-270mmHg
    Measurement Type: Systolic, Diastolic, Mean
    Resolution: 1mmHg
    NIBP Accuracy: Mean ±5mmHg, SD ±8mmHg
    Alarm: SYS/DIA/MEAN

    Respiration:
    Method: RA-LL impedance
    Measurement range: Adult: 0-120rpm, Paediatric/Neonatal: 0-150rpm
    Apnea Alarm: YES

    SpO2:
    Measurement range: 0%-100%
    Accuracy: ±2% (70%-100%); 0%-69% unspecified.
    Alarm range: 0-100%

    Pulse Rate:
    Measurement range: 10 – 270
    Resolution: 1bpm
    Accuracy: ±2pm or ±2%
    Temperature:

    Temprature:
    Resolution: 0.2 0C
    Channel: Dual-channel
    Measurement range: 0-50 0C

    Net Weight: 3.9kg
    Content of the box:
    Monitor
    Power connector
    Adult/Neonatal cuff + cable (based on selection)
    ECG leads
    SpO2 probe
    Temperature probe